Maintenance contracts for BOECO PCR Workstation with UV Air Recirculator
A service visit on the BOECO PCR Workstation with UV Air Recirculator covers airflow measurement — inflow and downflow velocities — HEPA filter integrity testing, alarm and interlock checks, and the lighting, UV lamp and electrical safety, following the standard the cabinet is certified to.
Every visit includes calibration and performance verification, and closes with a documented maintenance report you can present during an accreditation audit. Spare-parts availability and a guaranteed response time are written into the contract, so a breakdown never leaves a department waiting for an answer.
Containment and clean-air cabinets like the BOECO PCR Workstation with UV Air Recirculator are tested at installation, after they are moved or have a filter changed, and at least once a year — the interval NSF/ANSI 49 and EN 12469 set for biosafety cabinets and the usual practice for clean benches and fume hoods.
- Inflow and downflow velocities measured against the rated values
- HEPA filter integrity (leak) test
- Airflow alarms, interlocks and sash checked
- Lighting, UV lamp and electrical safety tested