Maintenance contracts for LabTech Contact Lens Sterilizer
A service visit on the LabTech Contact Lens Sterilizer covers calibration of the temperature and pressure sensors, the door seal, lock and safety valve, the steam generator, vacuum system, drains and filters, and a recorded test cycle.
Every visit includes calibration and performance verification, and closes with a documented maintenance report you can present during an accreditation audit. Spare-parts availability and a guaranteed response time are written into the contract, so a breakdown never leaves a department waiting for an answer.
Steam sterilisers like the LabTech Contact Lens Sterilizer are normally serviced once or twice a year and revalidated every year, while users run the routine tests their class requires — for Class B units, a daily steam-penetration (Bowie-Dick or Helix) test and a regular vacuum leak test.
- Temperature and pressure sensors calibrated
- Door seal, locking mechanism and safety valve checked
- Steam generator descaled, filters and drains cleaned
- Test cycles run and recorded, including vacuum leak and Bowie-Dick tests on Class B models